High-Resolution Transducers up to 1,000 MHz

With a frequency of 1,000 MHz, our transducers offer the world’s highest applied frequency for acoustic mapping and analysis. This means that resolutions down to 0,8 µm  µm can be achieved on suitable samples. PVA TePla manufactures transducers of 100–1,000 MHz for the most diverse high-resolution applications, based on sapphire lens technology and ZnO piezo materials of the active element.

These special lenses were developed in particular for the analysis of thin layers, but also for quantitative acoustic microscopy. In this case, Rayleigh waves are induced just below the surface of the material and analyzed using v(z) and f(v) curves. By means of suitable Fourier analysis, surface wave velocities (Rayleigh waves) can be measured with an accuracy exceeding 0.1%, which enables local determination of the modulus of elasticity or shear modulus.

Contact us

How can we help you?
Let´s talk!

sales-sg(at)pvatepla.com

PVA TePla Transducer Development

Generation of the electrical matching network to ensure optimum adaptation to the necessary application.

PVA TePla Transducer

PVA TePla Analytical Systems Transducer
 

Ultrasonic System

PVA TePla Analytical Systems Ultrasonic System Transducer

Transducer / Range

The table shows a list of available transducers. Additional versions are available on request and can be manufactured for specific applications.

No. Frequency [MHz] Axial resolution [µm] Detection limit [µm] Focal length in water [mm] Additionally available focal lengths (in water) [mm] Aperture diameter [mm]
1 Transducer 5 MHz 257.18 85.73 37.00 20 mm – 60 mm (customer-specific), or non-focused 3.5–6.36
2 Transducer 10 MHz 160.74 53.58 19.00 20–50 mm (customer-specific), or non-focused 3.5–6.36
3 Transducer 15 MHz 98.91 32.97 37.00 20–50 mm (customer-specific), or non-focused 3.5–6.36
4 Transducer 20 MHz 75.64 25.21 10.00 20–50 mm (customer-specific), or non-focused 2.5–5
5 Transducer 25 MHz 58.45 19.48 25.00 25 mm (customer-specific), or non-focused 2.5–5
6 Transducer 30 MHz 49.46 16.49 13.00 12.7 mm 4
7 Transducer 35 MHz 42.86 14.29 30.00 30 mm 4
8 Transducer 35 MHz 42.86 14.29 35.00 40 mm 4
9 Transducer 50 MHz 29.90 9.97 12.00 10–12 mm 3.0–4.0
10 Transducer 75 MHz 20.09 6.70 12.70 8–12.7 mm 3
11 Transducer 80 MHz 19.78 6.59 10.00 8–12 mm 3
12 Transducer 100 MHz 14.95 4.98 8.00 8 mm or 10 mm 2
13 Transducer 110 MHz 13.68 4.56 8.00 8 mm 2
14 Transducer 125 MHz 11.91 3.97 9.00 7–9 mm 2
15 Transducer 150 MHz 9.97 3.32 5.90 5.9 mm 1.85–2.2
16 Transducer 150 MHz 9.97 3.32 8.00 8 mm 1.85–2.2
17 Transducer 150 MHz 9.97 3.32 4.00 4 mm 1.85–2.2
18 Transducer 150 MHz 9.97 3.32 3.20 3.2 mm 1.85–2.2
19 Transducer 150 MHz 9.97 3.32 8.0 8 mm 1.85–2.2
20 Transducer 175 MHz 8.57 2.86 4.00 4 mm 1.85–2.2
21 Transducer 175 MHz 8.57 2.86 3.20 3.2 mm 1.85–2.2
22 Transducer 175 MHz 8.57 2.86 8.00 8 mm 1.85–2.2
23 Transducer 200 MHz 7.52 2.51 8.00 8 mm 1.85–2.2
24 Transducer 200 MHz 7.52 2.51 4.00 4 mm 1.85–2.2
25 Transducer 200 MHz 7.52 2.51 8.00 8 mm 1.85–2.2
26 Transducer 200–250 MHz 6.01 2.00 5.00 5 mm 1.85–2.2
27 Transducer 200–250 MHz 6.01 2.00 4.00 4 mm 1.85–2.2
28 Transducer 200–250 MHz 6.01 2.00 3.20 3.2 mm 1.85–2.2
29 Transducer 200–250 MHz 6.01 2.00 8.00 8 mm 1.85–2.2
30 Transducer 300 MHz 5.02 1.67 3.20 3.2 mm 1.85–2.2
31 Rayleigh wave lens 100 MHz 7.56 2.52 1.80 1.8 mm 1.8
32 Rayleigh wave lens 200 MHz 3.76 1.25 0.60 0.6 mm 0.6
32 Rayleigh wave lens 400 MHz 1.88 0.63 0.30 0.3 mm 0.3
33 Rayleigh wave lens 1,000 MHz 0.60 2.52 0.08 0.08 mm 0.08
34 Rayleigh wave lens 800–1,300 MHz 0.60 1.25 0.06 0.06 mm 0.06
35 Rayleigh wave lens 1,300–2,000 MHz 0.60 0.63 0.06 0.06 mm 0.06

Immediate Contact

slideoutcontact
Please complete all fields marked with *.
Please solve the following question:
captcha

PVA TePla Singapore Pte. Ltd.
12 New Industrial Road, #05-03
Morningstar Centre
Singapore, 536202

Telephone: +65 9631 4005